Back To Search Results

Hounsfield Unit

Editor: Johanna Schubert Updated: 3/6/2023 2:39:36 PM

Definition/Introduction

The Hounsfield unit (HU) is a relative quantitative measurement of radio density used by radiologists in the interpretation of computed tomography (CT) images. The absorption/attenuation coefficient of radiation within a tissue is used during CT reconstruction to produce a grayscale image. The physical density of tissue is proportional to the absorption/attenuation of the X-ray beam. The Hounsfield unit, also referred to as the CT unit, is then calculated based on a linear transformation of the baseline linear attenuation coefficient of the X-ray beam, where distilled water (at standard temperature and pressure) is arbitrarily defined to be zero Hounsfield Units and air defined as -1000 HU. The upper limits can reach up to 1000 for bones, 2000 for dense bones like the cochlea, and more than 3000 for metals like steel or silver. The linear transformation produces a Hounsfield scale that displays as gray tones. More dense tissue, with greater X-ray beam absorption, has positive values and appears bright; less dense tissue, with less X-ray beam absorption, has negative values and appears dark. The Hounsfield unit was named after Sir Godfrey Hounsfield, recipient of the Nobel Prize in Physiology or Medicine in 1979, for his part in the invention of CT, as it had immediate recognition as a revolutionary diagnostic instrument.[1][2][3]

Issues of Concern

Register For Free And Read The Full Article
Get the answers you need instantly with the StatPearls Clinical Decision Support tool. StatPearls spent the last decade developing the largest and most updated Point-of Care resource ever developed. Earn CME/CE by searching and reading articles.
  • Dropdown arrow Search engine and full access to all medical articles
  • Dropdown arrow 10 free questions in your specialty
  • Dropdown arrow Free CME/CE Activities
  • Dropdown arrow Free daily question in your email
  • Dropdown arrow Save favorite articles to your dashboard
  • Dropdown arrow Emails offering discounts

Learn more about a Subscription to StatPearls Point-of-Care

Issues of Concern

This linear transformation of the original linear attenuation makes the Hounsfield scale a relative scale, rather than absolute. Different X-ray beam energies will result in different tissue absorption and hence, different HUs. Early studies showed the HU to be dependent on the various CT parameters.[4] The type of reconstructing algorithm, design of the CT, and X-ray kilovoltage, were the most important factors identified. These factors need standardization to help make the HU a reliable diagnostic measurement tool.[5]

CT artifacts can affect Hounsfield unit measurements.  One of the most encountered CT artifacts, beam-hardening artifact, affects the measurement of radiodensity. Polychromatic energies comprise the conventional CT X-ray. High-density tissue selectively absorbs X-rays of lower energy, thus altering the X-ray beam. This absorption, in turn, can alter the X-ray beam absorption in the center of high-density tissue and results in a change in the HU, leading to falsely lowered HU measurements and appears less dense, or darker, on CT images. Modern CT machines can correct this artifact in the reconstruction process.

Continued advancements in CT as a diagnostic tool have led to different CT designs. Different CT designs, in turn, can alter the HU. For example, cone-beam computed tomography (CBCT), used mainly in dentistry, cannot show the actual HU similar to conventional CT but does show a strong correlation.[6] Dual-energy CT (DECT) uses X-ray beams of two different energies for deriving additional information to produce both anatomic and functional information.[7] Given the dependence of the HU on energy, the use of HU as a quantitative diagnostic parameter is limited in DECT.  The same is true for reduced energy protocols used in CT imaging today.[8]

Lastly, one should remember that the visualization of images on a CT has as its basis differences in tissue density and radiodensity. In the case of foreign body evaluation on CT imaging, if the foreign body has a similar physical density to the tissue that it is embedded in, it will have similar HU and will be hard to detect by visually CT. The radiologic evaluation of a wooden foreign body is complicated, given the varied appearance of wood and changes within the wood. There is documentation showing that a wooden foreign body demonstrates increasing HU over time.[9]

HU for fat is around -50, cerebrospinal fluid +15, white matter +25, grey matter +40, and blood +30-45.

Clinical Significance

The use of the HU to measure tissue density has aided radiologists in the interpretation of images and diagnosis of disease. Its use is identified in different specialties of medicine.

The following are some of the uses of HU that have been identified in the literature:

  1. In diagnosing fatty liver[10]
  2. As a quantitative measurement in the evaluation of solitary pulmonary nodules and thyroid nodules[11]
  3. To determine bone mineral density[12][13] 
  4. To estimate bone quality before spinal instrumentation,[14] predicting pedicle screw loosening[15] and in degenerative lumbar scoliosis[16]
  5. HU of paraspinal muscles have helped identify patients at risk for sarcopenia[17]
  6. In the characterization of bile duct lesions[18]
  7. Predictor of growth in meningiomas[19]
  8. Predictor of spontaneous expulsion of lower ureteric stone[20] and in differentiating pyonephrosis from hydronephrosis[21]
  9. HU of interpeduncular cistern hematomas can predict symptomatic vasospasms[22]
  10. In the differential diagnosis of odontogenic cysts[23]
  11. In predicting outcomes of intracerebral hemorrhage by measuring the HU in the perihematomal edema[24]
  12. In identifying tandem occlusion in acute ischemic stroke[25]
  13. In the diagnosis of acute venous thrombosis in the pediatric population[26]

With continued further advancements in technology, researchers are studying semi-automated measurements of tissue to aid the radiologist in the evaluation and diagnosis of disease. Semi-automated HU measurements of solitary pulmonary nodules have shown to be an accurate approach to determining malignant from benign solitary pulmonary nodules.[27] Other semi-automated measurements of HU are likely to aid in the evaluation of CT images and become part of the landscape of clinical practice in the near future.

Nursing, Allied Health, and Interprofessional Team Interventions

Proper communication should be there between the radiologist, surgeon, and physician with regard to imaging. Subtle findings may need the radiologist to measure the Housfiled unit of the lesion and let the medical or surgical team know if indicated.

References


[1]

Raju TN. The Nobel chronicles. 1979: Allan MacLeod Cormack (b 1924); and Sir Godfrey Newbold Hounsfield (b 1919). Lancet (London, England). 1999 Nov 6:354(9190):1653     [PubMed PMID: 10560712]


[2]

Mahesh M. Search for isotropic resolution in CT from conventional through multiple-row detector. Radiographics : a review publication of the Radiological Society of North America, Inc. 2002 Jul-Aug:22(4):949-62     [PubMed PMID: 12110725]


[3]

Hounsfield GN. Computed medical imaging. Nobel lecture, Decemberr 8, 1979. Journal of computer assisted tomography. 1980 Oct:4(5):665-74     [PubMed PMID: 6997341]

Level 3 (low-level) evidence

[4]

Levi C, Gray JE, McCullough EC, Hattery RR. The unreliability of CT numbers as absolute values. AJR. American journal of roentgenology. 1982 Sep:139(3):443-7     [PubMed PMID: 6981306]


[5]

Zerhouni EA, Spivey JF, Morgan RH, Leo FP, Stitik FP, Siegelman SS. Factors influencing quantitative CT measurements of solitary pulmonary nodules. Journal of computer assisted tomography. 1982 Dec:6(6):1075-87     [PubMed PMID: 7174924]


[6]

Razi T, Niknami M, Alavi Ghazani F. Relationship between Hounsfield Unit in CT Scan and Gray Scale in CBCT. Journal of dental research, dental clinics, dental prospects. 2014 Spring:8(2):107-10. doi: 10.5681/joddd.2014.019. Epub 2014 Jun 11     [PubMed PMID: 25093055]


[7]

Aran S, Daftari Besheli L, Karcaaltincaba M, Gupta R, Flores EJ, Abujudeh HH. Applications of dual-energy CT in emergency radiology. AJR. American journal of roentgenology. 2014 Apr:202(4):W314-24. doi: 10.2214/AJR.13.11682. Epub     [PubMed PMID: 24660729]


[8]

Bolus D, Morgan D, Berland L. Effective use of the Hounsfield unit in the age of variable energy CT. Abdominal radiology (New York). 2017 Mar:42(3):766-771. doi: 10.1007/s00261-017-1052-4. Epub     [PubMed PMID: 28132073]


[9]

Natung T, Shullai W, Lynser D, Tripathy T. A challenging case of a large intraorbital foreign body perforating the nasal septum in a child. Indian journal of ophthalmology. 2018 Oct:66(10):1511-1513. doi: 10.4103/ijo.IJO_150_18. Epub     [PubMed PMID: 30249858]

Level 3 (low-level) evidence

[10]

Zeb I, Li D, Nasir K, Katz R, Larijani VN, Budoff MJ. Computed tomography scans in the evaluation of fatty liver disease in a population based study: the multi-ethnic study of atherosclerosis. Academic radiology. 2012 Jul:19(7):811-8. doi: 10.1016/j.acra.2012.02.022. Epub 2012 Apr 21     [PubMed PMID: 22521729]


[11]

Wei PY, Jiang ND, Xiang JJ, Xu CK, Ding JW, Wang HB, Luo DC, Han ZJ. Hounsfield Unit Values in ACR TI-RADS 4-5 Thyroid Nodules with Coarse Calcifications: An Important Imaging Feature Helpful for Diagnosis. Cancer management and research. 2020:12():2711-2717. doi: 10.2147/CMAR.S242524. Epub 2020 Apr 22     [PubMed PMID: 32368148]


[12]

Scheyerer MJ, Ullrich B, Osterhoff G, Spiegl UA, Schnake KJ, Arbeitsgruppe Osteoporotische Frakturen der Sektion Wirbelsäule der Deutschen Gesellschaft für Orthopädie und Unfallchirurgie. [Hounsfield units as a measure of bone density-applications in spine surgery]. Der Unfallchirurg. 2019 Aug:122(8):654-661. doi: 10.1007/s00113-019-0658-0. Epub     [PubMed PMID: 31053924]


[13]

Narayanan A, Cai A, Xi Y, Maalouf NM, Rubin C, Chhabra A. CT bone density analysis of low-impact proximal femur fractures using Hounsfield units. Clinical imaging. 2019 Sep-Oct:57():15-20. doi: 10.1016/j.clinimag.2019.04.009. Epub 2019 May 8     [PubMed PMID: 31102777]


[14]

Zaidi Q, Danisa OA, Cheng W. Measurement Techniques and Utility of Hounsfield Unit Values for Assessment of Bone Quality Prior to Spinal Instrumentation: A Review of Current Literature. Spine. 2019 Feb 15:44(4):E239-E244. doi: 10.1097/BRS.0000000000002813. Epub     [PubMed PMID: 30063528]

Level 2 (mid-level) evidence

[15]

Zou D, Muheremu A, Sun Z, Zhong W, Jiang S, Li W. Computed tomography Hounsfield unit-based prediction of pedicle screw loosening after surgery for degenerative lumbar spine disease. Journal of neurosurgery. Spine. 2020 Jan 3:():1-6. doi: 10.3171/2019.11.SPINE19868. Epub 2020 Jan 3     [PubMed PMID: 31899883]


[16]

Wang H, Zou D, Sun Z, Wang L, Ding W, Li W. Hounsfield Unit for Assessing Vertebral Bone Quality and Asymmetrical Vertebral Degeneration in Degenerative Lumbar Scoliosis. Spine. 2020 Nov 15:45(22):1559-1566. doi: 10.1097/BRS.0000000000003639. Epub     [PubMed PMID: 32756284]

Level 2 (mid-level) evidence

[17]

Barnard R, Tan J, Roller B, Chiles C, Weaver AA, Boutin RD, Kritchevsky SB, Lenchik L. Machine Learning for Automatic Paraspinous Muscle Area and Attenuation Measures on Low-Dose Chest CT Scans. Academic radiology. 2019 Dec:26(12):1686-1694. doi: 10.1016/j.acra.2019.06.017. Epub 2019 Jul 17     [PubMed PMID: 31326311]


[18]

Batur A, Kerimoglu U, Ataseven H. Hounsfield unit density in the characterisation of bile duct lesions. Polish journal of radiology. 2019:84():e397-e401. doi: 10.5114/pjr.2019.89390. Epub 2019 Oct 9     [PubMed PMID: 31969956]


[19]

Nakasu S, Onishi T, Kitahara S, Oowaki H, Matsumura KI. CT Hounsfield Unit Is a Good Predictor of Growth in Meningiomas. Neurologia medico-chirurgica. 2019 Feb 15:59(2):54-62. doi: 10.2176/nmc.oa.2018-0209. Epub 2019 Jan 26     [PubMed PMID: 30686812]


[20]

Bokka S, Jain A. Hounsfield unit and its correlation with spontaneous expulsion of lower ureteric stone. Therapeutic advances in urology. 2019 Jan-Dec:11():1756287219887661. doi: 10.1177/1756287219887661. Epub 2019 Dec 3     [PubMed PMID: 31832102]

Level 3 (low-level) evidence

[21]

Boeri L, Fulgheri I, Palmisano F, Lievore E, Lorusso V, Ripa F, D'Amico M, Spinelli MG, Salonia A, Carrafiello G, Montanari E. Hounsfield unit attenuation value can differentiate pyonephrosis from hydronephrosis and predict septic complications in patients with obstructive uropathy. Scientific reports. 2020 Oct 29:10(1):18546. doi: 10.1038/s41598-020-75672-8. Epub 2020 Oct 29     [PubMed PMID: 33122830]


[22]

Ishihara H, Oka F, Kawano R, Shinoyama M, Nishimoto T, Kudomi S, Suzuki M. Hounsfield Unit Value of Interpeduncular Cistern Hematomas Can Predict Symptomatic Vasospasm. Stroke. 2020 Jan:51(1):143-148. doi: 10.1161/STROKEAHA.119.026962. Epub 2019 Nov 7     [PubMed PMID: 31694506]


[23]

Uehara K, Hisatomi M, Munhoz L, Kawazu T, Yanagi Y, Okada S, Takeshita Y, Saito EA, Asaumi J. Assessment of Hounsfield unit in the differential diagnosis of odontogenic cysts. Dento maxillo facial radiology. 2021 Feb 1:50(2):20200188. doi: 10.1259/dmfr.20200188. Epub 2020 Aug 12     [PubMed PMID: 32783633]


[24]

Huan R, Li Y, Tan J, Tang J, Huang N, Cheng Y. The Hounsfield Unit of Perihematomal Edema Is Associated With Poor Clinical Outcomes in Intracerebral Hemorrhage. World neurosurgery. 2021 Feb:146():e829-e836. doi: 10.1016/j.wneu.2020.11.025. Epub 2020 Nov 12     [PubMed PMID: 33189917]

Level 2 (mid-level) evidence

[25]

Mühl-Benninghaus R, Dressler J, Haußmann A, Simgen A, Reith W, Yilmaz U. Utility of Hounsfield unit in the diagnosis of tandem occlusion in acute ischemic stroke. Neurological sciences : official journal of the Italian Neurological Society and of the Italian Society of Clinical Neurophysiology. 2021 Jun:42(6):2391-2396. doi: 10.1007/s10072-020-04798-4. Epub 2020 Oct 14     [PubMed PMID: 33052575]


[26]

de la Vega Muns G, Quencer R, Ezuddin NS, Saigal G. Utility of Hounsfield unit and hematocrit values in the diagnosis of acute venous sinus thrombosis in unenhanced brain CTs in the pediatric population. Pediatric radiology. 2019 Feb:49(2):234-239. doi: 10.1007/s00247-018-4273-y. Epub 2018 Oct 16     [PubMed PMID: 30327829]


[27]

Choi Y, Gil BM, Chung MH, Yoo WJ, Jung NY, Kim YH, Kwon SS, Kim J. Comparing attenuations of malignant and benign solitary pulmonary nodule using semi-automated region of interest selection on contrast-enhanced CT. Journal of thoracic disease. 2019 Jun:11(6):2392-2401. doi: 10.21037/jtd.2019.05.56. Epub     [PubMed PMID: 31372276]